Daily Cone Beam CT QA
The QUASAR™ Penta-Guide phantom is used for the commissioning and daily testing of Image-Guided Radiotherapy (IGRT) systems. The Penta-Guide ensures the accuracy of linac-mounted image guidance systems, including cone beam CT (CBCT), x-ray volumetric imaging (XVI) and on-board imaging (OBI). Using low-density objects to minimize imaging artefacts, the Penta-Guide allows rapid and easy daily testing for:
- 3D cone beam registration
- kV and MV system coincidence
- kV and MV projection images
- Laser and light field coincidence
- Remote table adjustments
Used by a therapist every morning, testing takes little or no extra time and can be
completed during normal daily equipment warm-up procedures. Ideally, one
Penta-Guide phantom should be used per linac allowing simultaneous system testing
and
preventing unnecessary start-up delays.

The Penta-Guide also comes with a Lehmann Yoke for confirmation of the source to surface distance (SSD). A protocol outlining this enhanced testing procedure is included with the User’s Guide.
(Photo provided by Joerg Lehmann) | An optional Hood is used to confirm that the Varian Frameless Array or other visual position indicator is aligned with the CBCT and OBI. |
|